EHF-U
Table-Top Electric Hydraulic Dynamic and Fatigue Testing Machine
Test force | 50, 100, 200 kN |
Test type | Tensile, bending & compression tests |
Drive | Electrical Hydraulic mechanical |
Type | Table-top appliances |
Industries | Dynamic |
For Full-Scale Fatigue and Endurance Testing of Structural Materials and Large Samples
By providing a T-slot surface plate at the bottom of a U-type loading frame with a top-mounted actuator, these systems allow installation of extra large components and parts. Due to the large testing space, dynamic testing and evaluation can be performed on a variety of samples, including full-size samples and structural members.
1. Top-Mounted Actuator
Supports dynamic and endurance testing of full-size samples, large components, and structural members.
2 Hydraulically lifting/lowering crosshead
3. ±0.5 % test force accuracy
Test force accuracy is guaranteed to within ±0.5 % of the indicated value.

1. High Rigidity and Large Testing Space
A T-slot surface plate makes it easy to secure samples. The size of the surface plate and length of the columns can be customized based on the size of samples.
2 High-accuracy column
3. Crosshead Drive System with Operating Error Prevention Mechanism
The hydraulic crosshead drive and hydraulic clamp can be operated intuitively using the handle.

Amplitude Characteristics (60 Hz)


50kN±25mm

100kN±25mm

200kN±25mm

50kN±50mm

100kN±50mm

200kN±50mm
* It is not possible to use standard configurations of models QF-70B or higher for high-frequency regions due to the servo valve characteristics. However, these models may be used for testing at high frequencies if the servo valve is changed, for example. Contact Shimadzu for more information.
- The above characteristic curves indicate the relation between half-amplitude and cycle speed, given sine wave motion at the rated load level.
- The above indicates the amplitude characteristics given a 60 Hz power supply. Characteristics with a 50 Hz power supply will be about 5/6 of indicated values.
- The above characteristics do not include the frame or load cell characteristics. Compensate for the influence of these factors to determine actual amplitude characteristics.
- The indicated characteristics values were calculated based on typical characteristics of the servo valve being used, which may result in a difference of about 10 % on the frequency axis.
- There may be limitations on testing frequencies, depending on jig, sample, or other characteristics.
- At the installation site, provide about 500 mm of space on all four sides of the system, in addition to the space requirements indicated above, to allow access for operation and maintenance.
- The drawing above indicates the dedicated space requirements. The shape and orientation of the hydraulic power supply unit may vary depending on its capacity.
- For a more detailed standard layout drawing, contact Shimadzu.
- The standard system configuration does not include the table, computer, or printer.